The DXI-N-11 Axion defibrillator monitor with automatic external defibrillation function is designed for resuscitation and electropulse treatment of acute and chronic cardiac arrhythmias, determination of oxygen saturation and hemoglobin blood pressure, as well as the day of external, transesophageal and endocardial cardiac pacing. The DKI-N-11 "Axion" defibrillator monitor is used in medical hospitals, cardiology clinics, to equip emergency medical equipment and emergencies. The device is certified to meet the requirements of Directive 93/42 / EEC. The 'AXION' defibrillator-monitor is a new device in the product range of Ltd. Concern 'AXION'. All technical characteristics of the DKI-N-11 'AXION' model are retained: the formation of physiologically optimal bipolar pulse energy at 360 J. External, endocardial, and transesophageal stimulation functions. Electrodes for children and adults. Energy accumulation time 200 J - 6 sec, 360 J - 10 sec. Operator voice support and device operation. All information is displayed on the TFT screen and recorded on the thermal printer. 220 V power supply is added. The storage battery is built-in. There is a replaceable memory card.
Features:
Continuous operation of the device in monitoring mode from AC 220 V
Continuous operating time of the device in monitoring mode:
of a replaceable battery for at least 3 hours
from the DC source (12–18) V for at least 168 hours
Weight: no more than 7 kg.
ECG channel - 3/6/12 ECG leads
NIBP channel: automatic blood pressure measurement in the range of 20 to 280 mm Hg. with an absolute error of ± 3 mm Hg.
Protection against maximum pressure, over 330 mm Hg.
SpO2 channel - SpO2 readings range from 0 to 100%
The SpO2 reading range is from 75 to 100% with an absolute error of ± 3%.
EX channel - external, endocardial, and transesophageal
Channel EX: works in modes: fixed Fixed, overdrive Overdrive, and demand "on demand".
Frequency setting from 40 to 250 pulses/min, pulse duration setting from 20 to 40 ms,
Adjust the current pulse from 10 to 180 mA for external stimulation.
Disposable electrodes for external stimulation.
Defibrillation pulse: biphasic, trapezoidal, asymmetric, with a negative and positive half-wave current ratio (0.5 ± 0.1)
Power output lock when the patient's body resistance is below 12 ohms and above 200 ohms
Automatic limitation of defibrillation current to a level of (30 ± 10) A with a patient body resistance of less than 25 ohms
Automatic stabilization of pulse output parameters based on the patient's thoracic resistance in the range of 25 to 175 ohms.
Voice and visual support of operator actions and device operation.
7″ color TFT display (800 × 640 pixels)
Built-in thermal printer.
Built-in battery charger
Defibrillation:
Pulse: physiologically optimal bipolar pulse with automatic correction of duration and shape based on of the patient's impedance.
Defibrillation pulse energy, J for adults 5/10/25/50/75/150/200/250/300/360 (additional function of power accumulation over 200J, with adult electrodes) for children 5/10/25/50/75/100/150 (power off at 150 J when operating in child mode)
Energy accumulation time, no more than, sec. 200 J 6
Number of defibrillation impulses of the
storage battery completamente loaded, no more than 200 J 70 360 J 40
Accumulated energy retention time Indication of remaining seconds, with 30
automatic restarts subsiguientes of the stored energy to the internal load,
no more than, second
ECG:
Two ECG receiver channels from defibrillation electrodes and cable
ECG separado . ECG monitoring via a cable of 4-lead ECG I .. .II, aVR ... aVF Monitoring ECG via a 10-lead ECG cable (optional) I ... III, aVR ... aVF, V1 ... V6 Sensitivity of the ECG channel, mm/mV 5, 10, 20
Image movement speed, mm / sec 12.5; 25;
Measurement range of 50 HR, bpm from 30 to 300
The absolute measurement error of HR, bpm ± 2
Paper width, mm 58
Possibility of grid printing at a slope of 1 mm
Scanning speed of presencia , mm / seg 12,5; 25; 50
Seguridad:
for BF type defibrillation electrodes with defibrillation protection for separate electrodes of the CF type monitor with defibrillation protection through the pressure control channel arterial tipo BF for the SpO2 control channel type BF for CF type stimulation channels Stop of energy in the resistance of the patient's body less than 12 ohms and greater than 200 ohms, as well as in case of open and closed electrodes.
In case of defibrillation failure, manual reset of stored energy to the internal load.
Display: Pantalla
LCD color TFT
Screen size 152x91mm
Diagonal 7 ''
Resolution 800 x 640 pixels
Monitor recovery time after defibrillation, no more than 6 seconds.
Display information:
predefined energy values, three ECG leads, the status of the replacement battery , the upper and lower alarm limits for heart rate and the current value, the energy accumulation process, the current date and time, mode, the applied current and the patient's chest resistance, photoplethysmogram, pulse frequency, SpO2 value, systolic and diastolic blood pressure, stimulation channel settings, duplication of voice support messages for operator actions, and operation of the recording device
Power supply:
When operated from a replaceable storage battery Charging time of the battery, at least 4 h
When operated from the DC network 12-20 V
When operated from the AC network 190-250V at 50-60Hz frequency
Continuous operation period when operating on AC network, at least 168h
Power supply from the mains, no more than 210 VA
Continuous operation period in monitoring mode when
Operated from the rechargeable storage battery, at least 3 h